General
Cominox Leaking Steam from Door area
Resolutions 1) Door Seal - Check door seal for any damages or dirt. Clean the door seal or replace if it is damaged 2) Door Adjustment - Make sure the sterilizer is cooled down, then open the door. Unscrew the ...
Cleaning your Cominox
Clean the sterilization chamber and the internal disk on the door each day. Use a cotton cloth soaked in alcohol or distilled water. During cleaning, it is advisable to wear protective gloves. If any scale is found, use stainless steel polishing ...
How to start a cycle on Cominox Sterilclave
STARTING the CYCLE From STAND BY, press START to run the cycle shown on the display. During the cycle, the display will show: • The current date and time • The cycle in progress • The phase of the cycle in progress • The temperature in the ...